Model 8585 Features:
Measures velocity, volume, temperature and humidity without ever changing probes.
Stores over 999 data points and can take continuous measurements in a datalogging mode
Prints in real-time to a lightweight MicroPrinter, or downloads stored data to a windows software application
Data logs on a variable time base for IAQ studies
Recalls useful statistics such as max., min., and average
Also measures wet bulb and dew point
Model 8585 Specifications:
TEMPERATURE
Range: 0 to 140°F (-17.8 to 60°C)
Accuracy: ±0.5°F (±0.3°C)
Resolution: 0.1°F (0.1°C)
RELATIVE HUMIDITY
Range: 0 to 95% RH
Accuracy: ±3% RH
Resolution: 0.1% RH
WET BULB TEMPERATURE
Range: 40 to 140°F (5 to 60°C)
Resolution: 0.1°F (0.1°C)
DEW POINT TEMPERATURE
Range: 2 to 120°F (-15 to 49°C)
Resolution: 0.1°F (0.1°C)
DATA LOGGING CAPABILITIES
Storage Range: Up to 1871 samples, up to 285 test IDs
Intervals: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, or 30 seconds; 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 30, or 60 minutes
Model 8570 Features:
Designed for ease of use and rapid checking of air flow conditions.
Measures velocity, volume and temperature
Automatically averages readings, saving time and tedious hand calculations
Easy-to-use - minimal key operation
Displays average of readings instantly with the press of a button
Model 8570 Specifications:
TEMPERATURE
Range: 0 to 200°F (-17.8 to 93.3°C)
Accuracy: ±0.5°F (±0.3°C)
Resolution: 0.1°F (0.1°C)
Common Features:
Use as IAQ instruments in office or industrial buildings, factories, and warehouses
Measure velocity at hoods or HEPA filters
Read temperature concurrently with flow
Shipped as a kit which includes thermo-anemometer; articulated, telescopic 42-inch probe; 4 C-size batteries; battery eliminator; owner's manual; carrying case and calibration data sheet.
Common Specifications:
VELOCITY
Range: 20 - 9999 fpm (0.1 - 50 m/s)
Accuracy: ±3% rdg. or ±3 fpm, whichever is greater
Resolution: 1 fpm (0.01 m/s), 5 fpm = 500 fpm (0.05 m/s = 5.00 m/s)
VOLUMETRIC FLOW RATE
Range: Actual range is a function of maximum velocity and duct size
Duct Dimensions: 1 to 250" in increments of 0.1" (1 to 635 cm in increments of 0.1 cm)
TIME CONSTANT
Intervals: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, or 30 seconds
DISPLAY
Primary Display: 4-digit LCD, 0.6" (15mm) digit height
Secondary Display: 3.5-digit LCD, 0.3" (8mm ) digit height
PHYSICAL
Power Supply: Four "C" size batteries
Battery Life: 24 hours average
Meter Dimensions: 3.8 x 7.5 x 2.35" (9.65 x 19 x 6 cm)
Probe Dimensions: Length: 43.25" (110 cm); Tip Diam.: 0.276" (7.01 mm); Base Diam.: 0.395" (10.03 mm);
Knuckle Diam.: 0.377" (9.44 mm)