Features:
Flowrate feature makes simple calculations of volumetric flow rate when the user inputs the duct shape and size, K factor or horn size
Velocity measurements are made from the thermal sensor or a Pitot tube
Automatic conversion between actual and standard velocity readings
Direct calculation of dew point and wet bulb temperature - no psychrometric chart needed (Model 8386 only)
Heat flow function calculates heat transferred after a heating or cooling element (Model 8386 only)
Stable digital display when measuring fluctuating flow
Backlit display is easy to read in poor lighting conditions
Switchable from Imperial to metric units via dip switch located on rear of unit (all units shipped reading Imperial units)
40-inch telescoping probe with etched length marks to make duct traverse measurements easier
Optional portable printer provides hard copy documentation of your measurements
DATALOGGING CAPABILITIES
Allows user to log 1394 samples with time and date stamp
Simultaneously records all parameters available on each model
Single point and continuous data logging modes to fit your application
Data can be reviewed on-screen, printed or downloaded to a computer spreadsheet program
Statistics function displays average, maximum and minimum values, and the number of recorded samples
Specifications:
Velocity from Thermal Sensors (all models)
Range: 0 to 9,999 fpm (0 to 50 m/s)
Accuracy: ±3.0% rdg. ±3 fpm (±.015 m/s), whichever is greater (= 30 fpm)
Resolution: 1 fpm (0.01 m/s)
Velocity from a Pitot Tube (Models 8385 & 8386)
Range: 250 to 15,500 fpm (1.27 to 78.7 m/s)
Accuracy: ±1.5% at 2,000 fpm (10.06 m/s)
Resolution: 1 fpm (0.01 m/s)
Volumetric Flowrate (all models)
Range: Function of max. air velocity, pressure, duct size, and K factor
Duct Size (all models)
Range: 1 to 250" in increments of 0.1" (1 to 635 cm in increments of 0.1 cm)
Static/Differential Pressure (Models 8385 & 8386)
Range: -5 to +15 in.H2O (-1245 to +3735 Pa, or -9.3 to +28 mmHg)
Accuracy: ±1% of reading ±0.005 in.H2O (±1 Pa or ±0.01 mmHg)
Resolution: 0.001 in.H2O (1 Pa, 0.01 mmHg)
Temperature
Range: 8384 & 8385: 0 to 200°F (-17.8 to 93.3°C)
Range: 8386: 14 to 140°F (-10 to 60°C)
Accuracy: ±0.5°F (±0.3°C)
Resolution: 0.1°F (0.1°C)
Relative Humidity (Model 8386 only)
Range: 0 to 95% RH
Accuracy: ±3% RH
Resolution: 0.1% RH
Wet Bulb Temperature (Model 8386 only)
Range: 40 to 140°F (5 to 60°C)
Resolution: 0.1°F (0.1°C)
Dewpoint (Model 8386 only)
Range: 5 to 120°F (-15 to 49°C)
Resolution: 0.1°F (0.1°C)
Heat Flow (Model 8386 only)
Range: Function of flow rate, temperature, relative humidity and barometric pressure
Measure Available: Sensible Heat Flow, Latent Heat Flow, Total Heat Flow, Sensible Heat Factor
Units Measured: BTU/h, kW
Logging Capability (all models)
Range: Up to 1394 samples and 275 test id's (one sample can contain up to all 11 measure
ment types)
Intervals: 2, 5, 10, 15, 20, 30, 60 sec., 2, 5, 10, 15, 20, 30, 60 min.
General (all models)
Time Constant: 1, 2, ,5, 10, 15, 20 sec. intervals
Power Supply: Four AA size batteries
Meter Dimensions: 4.2 x 7.2 x 1.5" (10.7 x 18.3 x 3.8 cm)
Probe Dimensions: 40" (101.6 cm) L x 0.276" (7.01 mm) dia.
Meter Weight: 1.2 lbs. (0.54 kg)